The Open Medical Imaging Journal

2012, 6 : 1-11
Published online 2012 April 6. DOI: 10.2174/1874347101206010001
Publisher ID: TOMIJ-6-1

Plausibility of Image Reconstruction Using a Proposed Flexible and Portable CT Scanner

Jeff Orchard , Hwa Young Kim and John T.W. Yeow
David R. Cheriton School of Computer Science, University of Waterloo, Waterloo, Ontario N2L 3G1 Canada.

ABSTRACT

The very hot and power-hungry x-ray filaments in today's computed tomography (CT) scanners constrain their design to be big and stationary. What if we built a CT scanner that could be deployed at the scene of a car accident to acquire tomographic images before moving the victim? Recent developments in nanotechnology have shown that carbon nanotubes can produce x-rays at room temperature, and with relatively low power needs. We propose a design for a portable and flexible CT scanner made up of an addressable array of tiny x-ray emitters and detectors. In this paper, we outline a basic design, propose a strategy for reconstruction, and demonstrate the feasibility of reconstruction using experiments on a software simulation of the flexible scanner. These simulations show that reconstruction quality is stable over a wide range of scanner geometries, while progressively larger errors in the scanner geometry induce progressively larger errors. We also raise a number of issues that still need to be overcome to build such a scanner.